(7) Cold Stage

Features Benefits
Active LN2-cooled sample block Permits analysis of otherwise volatile materials in a vacuum
2 separate cooling circuits Permits a cold plate near the sample to be held at lower temperature than sample
On-stub heater cartridge option Additional fine control of sample temperature and also above ambient heating to ~100°C
On-stub thermometry Precise temperature measurement
Very stiff mechanical design Permits high resolution imaging (minimise drift during data acquisition)
X, Y, Z motion stage: ±12mm on all axes Full analytical coverage of the 22mm x 22mm sample stub
Pneumatic sample clamp Ensures good thermal contact (<80 K on stub). Released for easy sample exchange
Option for digital micrometer readouts Precision positioning

Custom rule
Photo of cold stage

Kore has designed a number of high-precision sample stages for operation in UHV environments. For matching the cryo preparation system, we have designed an X, Y, Z stage with ±12 mm of motion in all directions. The stage fits on to a 6" flange and is designed to maintain samples cold for surface analysis of otherwise volatile materials. Using on-stub thermometry and control of the LN2 flow, the sample can be held at any temperature from −180°C to room temperature to within ±1°C. The X, Y and Z motions are achieved through manual, rotary motion, and an option exists to equip the drives with digital micrometer readouts for greater positioning precision.

If you would like to discuss custom requirements for stages please contact sales who will be happy to discuss your specific needs.